Improved Edited HMBC for Bruker
Pulse program
-
Multiplicity edited HMBC with third order low-pass J-filter:
edhmbc.2
AU-programs for processing
-
Simple processing macro that splits, combines and processes the spectra
edhmbc2proc
Procedure for pulseprogram edhmbc
-
Start with a HMBC parameterset rpar HMBCGP all
-
Change pulse program to
edhmbc
pulprog edhmbc
-
Set
FnMODE to Echo-Antiecho
-
Set up the acquisition parameters
ased
-
Set
cnst6 to 125 Hz and
cnst8 to 165 Hz.
The preparation delay, d6
can initially be set to 65 ms.
-
Set TD for F1 to 4 times the desired digital resolution in the final spectra.
-
Set the number of scans to ~ 4× that of a decent HSQC
on the same sample. (The minimum number of scans is 4.) Keep the same receiver gain.
|