Carlsberg Laboratory
NMR center
Pulse sequence library
Coupling Constants in Proteins
TROSY Experiments
Hydrogen Bond Experiments
Small Molecules
Pulse Calibration
Disclaimer
|
J-HMBC and XLOC for Bruker
Download package:
For SGI:
gjhmbc.tar.gz
For Windows:
gjhmbc_b1.exe
Content of J-HMBC / XLOC package
| Pulse Programs: J-HMBC
|
axgjhmbclp1_21
|
AVANCE Version for use with XWinNMR2.x, 1st order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axgjhmbclp2_21
|
AVANCE Version for use with XWinNMR2.x, 2nd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axgjhmbclp3_21
|
AVANCE Version for use with XWinNMR2.x, 3rd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axgjhmbclp1
|
AVANCE Version for use with XWinNMR2.6, 1st order
LPJF, requires HMBC gradient program
|
|
|
axgjhmbclp2
|
AVANCE Version for use with XWinNMR2.6, 2nd order
LPJF, requires HMBC gradient program
|
|
|
axjhmbclp3
|
AVANCE Version for use with XWinNMR2.6, 3rd order
LPJF, requires HMBC gradient program
|
| Pulse Programs: Constant time J-HMBC
|
axctgjhmbclp1_21
|
AVANCE Version for use with XWinNMR2.x, 1st order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axctgjhmbclp2_21
|
AVANCE Version for use with XWinNMR2.x, 2nd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axctgjhmbclp3_21
|
AVANCE Version for use with XWinNMR2.x, 3rd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axctgjhmbclp1
|
AVANCE Version for use with XWinNMR2.6, 1st order
LPJF, requires HMBC gradient program
|
|
|
axctgjhmbclp2
|
AVANCE Version for use with XWinNMR2.6, 2nd order
LPJF, requires HMBC gradient program
|
|
|
axctjhmbclp3
|
AVANCE Version for use with XWinNMR2.6, 3rd order
LPJF, requires HMBC gradient program
|
| Pulse Programs: Broadband XLOC
|
axbbxloclp1_21
|
AVANCE Version for use with XWinNMR2.x, 1st order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axbbxloclp2_21
|
AVANCE Version for use with XWinNMR2.x, 2nd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axbbxloclp3_21
|
AVANCE Version for use with XWinNMR2.x, 3rd order
LPJF, requires 2sineea2ns.r gradient program
|
|
|
axbbxloclp1
|
AVANCE Version for use with XWinNMR2.6, 1st order
LPJF, requires HMBC gradient program
|
|
|
axbbxloclp2
|
AVANCE Version for use with XWinNMR2.6, 2nd order
LPJF, requires HMBC gradient program
|
|
|
axbbxloclp3
|
AVANCE Version for use with XWinNMR2.6, 3rd order
LPJF, requires HMBC gradient program
|
|
Gradient Programs: |
4sineea34.r
|
Gradient program to be used with axgjhmbclp1_21 pulse
programs
|
|
|
5sineea45.r
|
Gradient program to be used with axgjhmbclp2_21 pulse
programs
|
|
|
6sineea56.r
|
Gradient program to be used with axgjhmbclp3_21 pulse
programs
|
|
|
HMBC
|
Gradient program to be used with axgjhmbclp>x<
pulse programs
|
| XWinNMR AU Programs: |
bbxloc
|
AU program for interactive setup and start of experiment)
|
|
|
au_bbxloc
|
AU BBXLOC acquisition program to be used in automation)
|
|
|
proc_gjhmbc
|
AU GJHMBC processing program to be used in automation
or in foreground
|
|
|
proc_bbxloc
|
AU BBXLOC processing program to be used in automation
or in foreground
|
|
Shell Scripts: |
gjhmbc_install
|
Script for installation of J-HMBC / XLOC package components
|
Programs tested on:
Bruker DRX600, XWinNMR 2.6 patchlevel 4, Irix 6.3
and Bruker DRX250, XWinNMR 2.1 patchlevel 7, Irix 5.3
Installation on SGI Workstations:
For installation the compressed archive
gjhmbc.tar.gz
containing the source codes of all components is required.
The installation is executed for all users on the system.
Content of installation package:
gjhmbc_b1.tar.gz, gjhmbc_install, GJHMBC_README
-
Save the
gjhmbc.tar.gz file at a location with write
permission for the user who is carrying out the installation.
-
Unpack the file with the commands:
gunzip gjhmbc.tar.gz
and
tar -xvf gjhmbc.tar
-
Run the
gjhmbc_install script,
answer all questions.
-
Installation is completed successfully when
the installation script runs without any error.
The installation files can be removed after successful installation.
-
Compile AU programs within XWinNMR prompt with
cpluser *gjhmbc *bbxloc
Installation on Windows NT:
The Windows NT version is provided as self extracting
archive gjhmbc_b1.exe.
-
Save the file on a disk that is accessible from the
computer that runs the spectrometer.
-
Login as user with write permission in the
\Bruker\Xwin-nmr directory.
-
Execute the
gjhmbc_b1.exe file.
-
Press the "
Browse" button
and choose the \Bruker\Xwin-nmr
directory in the "unzip to folder" window.
-
Compile AU programs within XWinNMR prompt with
cpluser *gjhmbc *bbxloc
Performing Experiments:
-
If no previous GJHMBC / XLOC dataset is available,
the parameters of a conventional HMBC can be used to start with.
-
Set the pulse sequence of choice in eda
-
Adjust the acquisition parameters according to the comments
in the pulse sequence (the delays
d21-d29 will
be evaluated when using the automation AU program au_jhmbc.
The number of non-zero values determine the number of subspectra
to be added in absolute-value mode.
In the interactive mode using the bbxloc AU program,
these values are ignored.
-
-
To start a XLOC experiment use the interactive
bbxloc AU program or the
au_bbxloc AU acquisition program
(automation setup see below).
-
GJHMBC can be with
zg
-
For processing run the
proc_gjhmbc/
proc_bbxloc AU program.
This will perform the complete processing,
and combination for Broadband XLOC automatically.
Notice that a zero-order phase correction in the F1 dimension
of 90 degrees is required (F1: PH_mod: pk, PHC0: 90),
to obtain absorption line shape in this dimension.
Automation:
GJ-HMBC and Broadband XLOC experiments
can easily be used in automation like ICONNMR.
The setup is similar to other SW optimized experiments.
-
Save an parameter set with
AUNM=au_getlinv and
AUNMP=proc_gjhmbc for GJHMBC
-
Save an parameter set with
AUNM=au_bbxloc and
AUNMP=proc_bbxloc
and appropriate values for
d21-d29 for BB XLOC.
-
Create a composite experiment within the setup menu of ICONNMR
and make it available for all users.
Notes:
-
When running BBXLOC in automation,
the experimental time is calculated wrong.
The real experimental time is
(Displayed_Value×Number_of_Delays).
-
Consecutive experimental numbers (
expno)
according to the number of broadband delays will be used.
I.e., If the BBXLOC-experiment is set up on experiment
number 10, then expnos 10, 11, 12 and 13
will be used (with four different delays).
IconNMR will not know this in advance and will propose
expno 11 for the next experiment.
You have to manually change this to 14
(or 20, 100 or whatever as long it is not in the range 11-13...).
-
|