MBOB for Bruker

Download package:

For SGI: mbob.tar.gz
For Windows: mbob_b1.exe

Content of MBOB package

Pulse Programs: axmboblp1_21 AVANCE Version for use with XWinNMR2.x, 1st order LPJF, requires 2sineea2ns.r gradient program
axmboblp2_21 AVANCE Version for use with XWinNMR2.x, 2nd order LPJF, requires 2sineea2ns.r gradient program
axmboblp3_21 AVANCE Version for use with XWinNMR2.x, 3rd order LPJF, requires 2sineea2ns.r gradient program
axmboblp1 AVANCE Version for use with XWinNMR2.6, 1st order LPJF, requires HMBC gradient program
axmboblp2 AVANCE Version for use with XWinNMR2.6, 2nd order LPJF, requires HMBC gradient program
axmboblp3 AVANCE Version for use with XWinNMR2.6, 3rd order LPJF, requires HMBC gradient program
axhmbclp1_21 AVANCE Version for use with XWinNMR2.x, 1st order LPJF, requires 4sineea34.r gradient program
axhmbclp2_21 AVANCE Version for use with XWinNMR2.x, 2nd order LPJF, requires 5sineea45.r gradient program
axhmbclp3_21 AVANCE Version for use with XWinNMR2.x, 3rd order LPJF, requires 6sineea56.r gradient program
axhmbclp1 AVANCE Version for use with XWinNMR2.6, 1st order LPJF, requires HMBC gradient program
axhmbclp2 AVANCE Version for use with XWinNMR2.6, 2nd order LPJF, requires HMBC gradient program
axhmbclp3 AVANCE Version for use with XWinNMR2.6, 3rd order LPJF, requires HMBC gradient program
Gradient Programs: 2sineea2ns.r Gradient program to be used with axmboblp>x<_21 pulse programs
4sineea34.r Gradient program to be used with axhmbclp1_21 pulse programs
5sineea45.r Gradient program to be used with axhmbclp2_21 pulse programs
6sineea56.r Gradient program to be used with axhmbclp3_21 pulse programs
HMBC Gradient program to be used with axmboblp>x< and axhmbclp>x< pulse programs
XWinNMR AU Programs: mbob AU program for interactive setup and start of experiment)
au_mbob AU MBOB acquisition program to be used in automation)
proc_mbob AU MBOB processing program to be used in automation or in foreground
bbhmbc AU Broadband HMBC acquisition program for interactive setup and start of experiment
au_bbhmbc AU Broadband HMBC acquisition program to be used in automation
proc_bbhmbc AU Broadband HMBC processing program to be used in automation or in foreground
Shell Scripts: mbob_install Script for installation of MBOB package components

Programs tested on:

Bruker DRX600, XWinNMR 2.6 patchlevel 4, Irix 6.3 and Bruker DRX250, XWinNMR 2.1 patchlevel 7, Irix 5.3

Installation on SGI Workstations:

For installation the compressed archive mbob.tar.gz containing the source codes of all components is required. The installation is executed for all users on the system.

Content of installation package: mbob_b1.tar.gz, mbob_install, MBOB_README

  1. Save the mbob.tar.gz file at a location with write permission for the user who is carrying out the installation.
  2. Unpack the file with the commands: gunzip mbob.tar.gz and tar -xvf mbob.tar
  3. Run the mbob_install script, answer all questions.
  4. Installation is completed successfully when the installation script runs without any error. The installation files can be removed after successful installation.
  5. Compile AU programs within XWinNMR prompt with cpluser *mbob

Installation on Windows NT:

The Windows NT version is provided as self extracting archive mbob_b1.exe.
  1. Save the file on a disk that is accessible from the computer that runs the spectrometer.
  2. Login as user with write permission in the \Bruker\Xwin-nmr directory.
  3. Execute the mbob_b1.exe file.
  4. Press the "Browse" button and choose the \Bruker\Xwin-nmr directory in the "unzip to folder" window.
  5. Compile AU programs within XWinNMR prompt with cpluser *mbob

Performing Experiments:

  1. If no previous MBOB dataset is available, the parameters of a conventional HMBC can be used to start with.
  2. Set the pulse sequence of choice in eda
  3. Adjust the acquisition parameters according to the comments in the pulse sequence (the delays d21-d29 will be evaluated when using the automation AU program au_mbob. The number of non-zero values determine the number of subspectra to be added in absolute-value mode. In the interactive mode using the mbob AU program, these values are ignored.
  4. To start the experiment use the interactive mbob AU program or the au_mbob AU acquisition program (automation setup see below)
  5. For processing run the proc_mbob AU program. This will perform the complete dataset editing, sorting, processing, and combination automatically. The edited one-bond correlation and multiple-bond correlation spectra will be plotted. Notice that a zero-order phase correction in the F1 dimension of 90 degrees is required (F1: PH_mod: pk, PHC0: 90), to obtain absorption line shape in this dimension.
    • The multiple-bond correlation spectrum is stored under the first expno and procno 100
    • The one-bond correlation spectrum is stored under the first expno and procno 200.

Automation:

MBOB experiments can easily be used in automation like ICONNMR. The setup is similar to other SW optimized experiments.
  1. Save an MBOB parameter set with AUNM=au_mbob and AUNMP=proc_mbob and appropriate values for d21-d29.
  2. Create a composite experiment within the setup menu of ICONNMR and make it available for all users.

Notes:

  • When running MBOB in automation, the experimental time is calculated wrong. The real experimental time is (Displayed_Value×4×Number_of_Delays).
  • Consecutive experimental numbers (expno) according to the number of broadband delays will be used. I.e., If the MBOB-experiment is set up on experiment number 10, then expnos 10, 11, 12 and 13 will be used (with four different delays). IconNMR will not know this in advance and will propose expno 11 for the next experiment. You have to manually change this to 14 (or 20, 100 or whatever as long it is not in the range 11-13...).
  • The MBOB-sequence has been tested under XWinNMR 3.5. However, some minor changes in the proc_mbob AU-program is necessary due to new plotting routines. Download the new proc_mbob to the AU-directory and the layout-file for XWinPLOT (MBOB.xwp) to the directory $XWINMRHOME/plot/layout.
 
Document: MBOB for Bruker (index.shtml)
Last modified: 2005-12-19