Sequential HNCACB

Ref.:A.Meissner, and O.W.Sørensen J.Magn.Reson. in press (2001). (Sequential HNCACB and CBCANH Protein NMR Pulse Sequences)

Sequential HNCACB pulse sequence
Delays: ' = S =  (2JNH)-1I = (2JNH)-1ICO = (2JCCO)-1 = gradient delay; TC = 14.3 ms, T = 22 ms, T' = 21 ms.
Phases for Varian Unity Inova:   = -y; {I = y; S = y - S } echo (t2 -> t3), {I = -y; S = -y - S } antiecho (t2 -> t3);
Phases for Bruker DRX:   = y; " = ; {I = -y; S = -y - S } echo (t2 -> t3), {I = y; S = y - S } antiecho (t2-> t3);
The uncompensated TROSY experiment employs SI = 0.
Gradient ratios: (G1 : G2 : G3) = (-7 : 3 : 1.987) echo; (G1 : G2 : G3) = (-8 : 2 : 3.013) antiecho.



Superposition of first t2 interferogram (echo) of sequential HNCACB (interresidual correlations) and edited intra residual HNCACB obtained by linear combination of HNCACB TROSY and sequential HNCACB in a ratio of 0.75 : -1. Sample: 4.18 mM 15N,13C-labeled CI2 21-83 (90% H2O/10% D2O, 25 deg. C, pH 4.2)


Varian:

exp s: sequential HNCACB (default)
c: HNCACB TROSY (shaped CO  pulses are omitted). 

To generate a selective water pulse shape, adjust ref_pwr and ref_pw90 in the file $HOME/vnmrsys/shapelib/water_sc.inp according to your proton pulse calibration and type the command: Pbox -f water_sc.inp

An output like
>>> Set pulse width to 1.0000 ms <<<
>>> Set pulse power to 21 dB <<<
appears.

Remember to change the parameters pwater (pulse width in us) and pwaterlvl (pulse power).

For generating a selective CO  pulse adjust ref_pwr and ref_pw90 in the file $HOME/vnmrsys/shapelib/ib_cob_off.inp according to your carbon pulse calibration. Adjust also the offset (ofs) to 140*dfrq when running at dof=36ppm. Apply the command: Pbox -f ib_cob_off.inp. change the parameters pwxcoilvl and pwxcoi according to the output. Apply the same procedure to the wurst2_cob.inp file.
 

Processing of 3D sequential HNCACB:

Data is States-TPPI in F1 and echo antiecho in F2. use your usual software for processing.

Installation:

For installation the compressed archive hncacb_seq.tar.Z containing all components is required. Copy the file into your $HOME directory uncomress (uncompress hncacb_seq.tar.Z) the file and unpack the archive (tar -xvf hncacb_seq.tar). The files will be copied to the appropriate location within the vnmrsys directory. Load the parameters within vnmr ($HOME/vnmrsys/parlib/hncacb_seq.par) and type the command seqgen(seqfil). Adjust Acquisition parameters (pulse lengths, offsets sweep widths, etc)

Program tested on: Varian Unity Inova, Vnmr6.1B, Pbox6.1C, Solaris7

Content
Pulse sequence:        hncacb_seq.c
Parameters:            hncacb_seq.par